Note:
Exposue.ppt
T. Toth 03-Aug-04
prescribed image coverage
exposure time = total x-ray on time
helical total coverage= (exposure time * table velocity)
axial or cine total coverage = macro row thickness * no of active rows
Slice thickness = nominal prospective reconstructed slice selection
image thickness
views to make image
exposure = (exposure time) * (image avg mA)* (slice thickness)/(total coverage)
Total coverage
mA
(avg mA for image)
Decsciption of how to calculate the Dicom Exposure field (0018,1152)
A.4.5 Frame of Reference Module
Attribute Name Tag Type Notes
Frame of Reference UID (0020,0052) 1 See Caution below.
Position Reference Indicator (0020,1040) 2
Value as entered at the user
interface.
A.4.6 Other Attributes
Attribute Name Tag Type Notes
Temporal Position Index (0020,9128) 3
Filled into CT image for some
applications. Pass number for
shuttle images.
Nominal Percentage of
Cardiac Phase
(0020,9241) 3
Filled into CT image for some
applications. Prescribed
percentage of cardiac phase.
It is possible for the operator of CT LightSpeed and BrightSpeed
system to change the table height while scanning a series of
images. Therefore, implementations must use the Frame of
Reference UID (0020,0052) in conjunction with the Table Height
(0018,1130) to determine if two images are spatially related.
CAUTION