GE H
EALTHCARE
D
IRECTION
DOC0500644, R
EV
4 L
IGHT
S
PEED
* DICOM C
ONFORMANCE
S
TATEMENT
111 of 157
• Referenced Patient Sequence – This sequence is not added
• Scheduled Protocol Code Sequence – The sequence is not send in the MPPS
message
• Performed Procedure Step Start date & Performed Procedure Step start time – The
exam date and exam time that is the Start date and Start Time of the associated Study
Component (Exam) is used, not the actual time when post-processing started.
• Performed Procedure Step end date & Performed Procedure Step end time - The date
and time when user chooses to ‘Complete PPS’ or ‘Discontinue PPS’ is taken as the
Performed Procedure Step End Date and Performed Procedure Step End Time
respectively (Not the actual end date and end time of post-processing).
• Procedure Code Sequence – This sequence is sent with ZERO items in the MPPS
message
• Performed Protocol Code Sequence – This sequence is sent with ZERO items in the
MPPS message.
• Referenced Standalone SOP Instance Sequence – The sequence is sent with ZERO
items in the MPPS message.
7.9 ERROR HANDLING AND RECOVERY
PPS Server AE does not define any extended error codes. The standard error codes are
handled. On a response with status ‘success’ for the N-CREATE or N-SET request, the
system updates the state and indicates the same on the user interface. On a response
with status other than ‘success’ the operation is deemed ‘Failed’ and the system updates
the state and indicates the same on the user interface. If the request has failed or
response is not received before the association timeout, the operation is deemed ‘Failed’
and the system updates the state and indicates the same on the user interface.
If the operation is ‘Failed’, detailed message is logged into system log-file and system
provides an alternative mechanism to retry the failed operation through the user interface
to ensure that transient failures do not affect the feature performance.