GE H
EALTHCARE
D
IRECTION
DOC0500644, R
EV
4 D
ISCOVERY
, L
IGHT
S
PEED AND
B
RIGHT
S
PEED
DICOM C
ONFORMANCE
S
TATEMENT
142 of 157
Performed Procedure
Step Start Date
(0040,0244) 3 Sent when MPPS option is enabled
Performed Procedure
Step Start Time
(0040,0245) 3 Sent when MPPS option is enabled
Performed Procedure
Step ID
(0040,0253) 3 Sent when MPPS option is enabled
Performed Procedure
Step Description
(0040,0254) 3 Sent when MPPS option is enabled
Requested Attributes
Sequence
(0040,0275) 3
Sent when MPPS option is enabled and
filled in based on what RIS provides to
scanner. Number of items relates to
number of items selected from Patient
Schedule.
> Requested
Procedure ID
(0040,1001) 1C Always sent, copied from RIS
> Accession Number (0008,0050) 3 Value from worklist sent
> Referenced Study
Sequence
(0008,1110) 3 Value from worklist sent
> Study Instance UID (0020,000D) 3 Value from worklist sent
> Requested
Procedure Desc
(0032,1060) 3 Value from worklist sent
> Requested
Procedure Code
Sequence
(0032,1064) 3 Value from worklist sent
> Scheduled
Procedure Step Desc
(0040,0007) 3 Value from worklist sent
> Scheduled Protocol
Code Sequence
(0040,0008) 3 Value from worklist sent
> Scheduled
Procedure Step ID
(0040,0009) 3 Value from worklist sent
> Reason for
Requested Procedure
(0040,1002) 3 Value from worklist sent
> Reason for
Requested Procedure
Code sequence
(0040,1004) 3 Value from worklist sent
A.3.5 General Equipment Module
Attribute Name Tag Type Notes
Manufacturer (0008,0070) 2 Always sent as “GE MEDICAL SYSTEMS”
Institution Name (0008,0080) 3 Sent. Value is configurable.
Station Name (0008,1010) 3 Sent. Value is configurable.
Manufacturers Model
Name
(0008,1090) 3
Device Serial
Number
(0018,1000) 3 Sent if present in image.